Initial clinical validation of a hybrid in silico—in vitro cardiorespiratory simulator for comprehensive testing of mechanical circulatory support systems

Fresiello, Libera and Muthiah, Kavitha and Goetschalckx, Kaatje and Hayward, Christopher and Rocchi, Maria and Bezy, Maxime and Pauls, Jo P. and Meyns, Bart and Donker, Dirk W. and Zieliński, Krzysztof (2022) Initial clinical validation of a hybrid in silico—in vitro cardiorespiratory simulator for comprehensive testing of mechanical circulatory support systems. Frontiers in Physiology, 13. ISSN 1664-042X

Full text not available from this repository.
Link to published document: http://doi.org/10.3389/fphys.2022.967449

Abstract

Simulators are expected to assume a prominent role in the process of design—development and testing of cardiovascular medical devices. For this purpose, simulators should capture the complexity of human cardiorespiratory physiology in a realistic way. High fidelity simulations of pathophysiology do not only allow to test the medical device itself, but also to advance practically relevant monitoring and control features while the device acts under realistic conditions. We propose a physiologically controlled cardiorespiratory simulator developed in a mixed in silico-in vitro simulation environment. As inherent to this approach, most of the physiological model complexity is implemented in silico while the in vitro system acts as an interface to connect a medical device. As case scenarios, severe heart failure was modeled, at rest and at exercise and as medical device a left ventricular assist device (LVAD) was connected to the simulator. As initial validation, the simulator output was compared against clinical data from chronic heart failure patients supported by an LVAD, that underwent different levels of exercise tests with concomitant increase in LVAD speed. Simulations were conducted reproducing the same protocol as applied in patients, in terms of exercise intensity and related LVAD speed titration. Results show that the simulator allows to capture the principal parameters of the main adaptative cardiovascular and respiratory processes within the human body occurring from rest to exercise. The simulated functional interaction with the LVAD is comparable to the one clinically observed concerning ventricular unloading, cardiac output, and pump flow. Overall, the proposed simulation system offers a high fidelity in silico-in vitro representation of the human cardiorespiratory pathophysiology. It can be used as a test bench to comprehensively analyze the performance of physically connected medical devices simulating clinically realistic, critical scenarios, thus aiding in the future the development of physiologically responding, patient-adjustable medical devices. Further validation studies will be conducted to assess the performance of the simulator in other pathophysiological conditions. Copyright © 2022 Fresiello, Muthiah, Goetschalckx, Hayward, Rocchi, Bezy, Pauls, Meyns, Donker and Zieliński.

Item Type: Article
Subjects: R Medicine > R Medicine (General)
Depositing User: Repository Administrator
Date Deposited: 02 Mar 2023 01:20
Last Modified: 02 Mar 2023 01:20
URI: http://eprints.victorchang.edu.au/id/eprint/1303

Actions (login required)

View Item View Item